Sunday, November 22, 2009
Cannon Lab 3T Scanning
    Notes
fMRI Logger
Safety Screen
Neurocognitive Screen


Compiled by DCG and NMH on 8/7/00

This checklist is to be used in conjunction with the Cannon Lab 3T Scanning Notes. It is not an explanation of how to scan!

Please enter the following information



Please enter the administrator's name:
Please enter the administrator's e-mail*:
Please enter subject ID:
Please enter Protocol/Condition:

Log into the fMRI Logger


Enter your password, EXPERIMENTER NAME, RESEARCH PROTOCOL & Enter the exam number
Enter Exam number from signa console
Enter the Signa ID (ID that is meaningful to the experimenter)
Do Safety and NeuroCog Screens
Record the HSPC# for the subject on the consent form and in the logger box.
  HSPC#:
  Problems with logger? Yes No
   
Consent the Subject

Subject completed metal screen and is currently metal free.
Subject has read and sign (in BOTH places) the STRUCTURAL & FUNCTIONAL MRI consent form
Leave the STRUCTURAL & FUNCTIONAL MRI form consent form in the signed consent bin.
HSPC# is written on the consent form.
  Problems with Consent? Yes No
   
Register the Subject on the Signa
Create Signa ID. Do not use spaces, "!", ";", or other fancy characters.
Enter: Signa ID, Subject Name, Sex, Age, Weight, Description of Scan.
Enter the signa ID on the logger.
  Signa ID: ; Exam Number:
  Problems with Registration? Yes No
   
Position the Subject in the Scanner
Have subject put in earplugs
Have subject put on headphones
Add lateral padding to reduce movement. Supplement with masking tape on the forehead.
Give subject the emergency alarm bulb.
Give subject the Goggles. Adjust the visual correction lenses as need.
Give subject the button box. Have the subject position the box on his leg.
Landmark and Position subject in scanner.
  Problems with Positioning the Subject? Yes No
   
Confirm that the Subject is O.K.
Call into subject.
Check equipment i.e. goggles
  Problems with Confirmation? Yes No
   
Run a Sagittal Scout
Setup Scan
Warn subject about noise.
Run Sag. Scout Scan
Exam scan for abnormalities. Contact Dr. Bookheimer or Dr. Cohen if necessary.
  Problems with Sagittal Scout? Yes No
   
Check that EPI Data will be Placed on Correct APD Disk
Check acquisition disk on ADP II, change if necessary.
Check acquisition disk on a X Window, change if necessary. ADP Disk:
  Problems with Acquisition Disk Check? Yes No
   
Shimming the Magnet
Select Shim/ShimmerSeq.
Review Screen
Make sure R/L direction is selected.
Run Shim/ShimmerSeq.
Run Shimmer on Instascan
  Problems with Shim? Yes No
   
Acquire a High-Resolution T2 Weighted Image
Select Basic Functional III- protocol specific
High Resolution Co-Planer- Oblique
Select AC-PC
Screen save
  Problems with High-Res. T2? Yes No
   
Acquire Functional EPI Images
Select an oblique EPI series
Erase slices if necessary
Set number of acquisition
Set scan timing: TR & TE
Set [water]
Set frequency to R/L to avoid shocking subject
Check matrix size (64 x128)
Record Scan Dimensions:No. of Slices: ; Matrix size: ;
Interscan Spacing:
Setup scan on signa and Instascan
Set up Mac stimulus delivery system
Run EPI at 10:24
   
Task Order
Task 1
Task 2
Task 3
Task 4
  Problems with Funct. EPI? Yes No
   
Acquire the T1 Structural Scan
Select msc 9minT1, (3D volume scan)
Select the coronal scout (8sec)
Choose 3D T1 (9 min)
Set and deposit scan box.
Scan
  Problems with T1 Structural Scan? Yes No
   
Liberate the Subject
Problems liberating the subject? Yes No
   
Extracting the Structural Scans
Problems with Structural Extraction? Yes No
   
Log out of the fMRI Logger
   
FTPing data from HBM
Call 5-9142 to see if anyone is on the scanner
FTP EPI images
FTP Structural Images
Remove Images from Signa
  Problems FTPing? Yes No
   
Comments: